Siemens Canada Siemens Milltronics Process Instruments LR560 SITRANS LR560 Level Probing Radar User Manual LT01 LR560PA
Siemens Canada Ltd. - Siemens Milltronics Process Instruments SITRANS LR560 Level Probing Radar LT01 LR560PA
Contents
User Manual PA











![Page 6 SITRANS LR560 (PROFIBUS PA) – OPERATING INSTRUCTIONS 7ML19985LT01mmmmmSITRANS LR560SITRANS LR560 OverviewSITRANS LR560 is a 2-wire, 78 GHz FMCW radar level transmitter for continuous monitoring of solids in vessels to a range of 100 m (329 ft). The plug and play performance is ideal for all solids applications, including those with extreme dust and high temperatures to +200 °C (+392 °F). The device consists of an electronic circuit coupled to a lens antenna and flange for quick and easy positioning. The main benefits of using 78 GHz over devices using lower frequency are:• very narrow beam, so device is insensitive to mounting nozzle interference and vessel obstructions.• short wavelength yields very good reflection properties on sloped solids, so aiming towards material angle of repose is usually not necessary.The technology is very tolerant of buildup on the lens antenna, however an air purge inlet is provided for periodic cleaning if required.SITRANS LR560 supports PROFIBUS PA communication protocol, and SIMATIC PDM software. Signals are processed using Process Intelligence which has been field-proven in over 1,000,000 applications worldwide (ultrasonic and radar). This device supports acyclic communications from both a PROFIBUS Class I and Class II master.ProgrammingSITRANS LR560 is very easy to install and configure via an optional graphical local display interface. You can modify the built-in parameters either locally via the push buttons or using the infrared handheld programmer, or from a remote location using one of the following options:• PROFIBUS PA (using SIMATIC PDM, FDT [such as PACTware or Fieldcare]) • HART (using 375 handheld Field Communicator, SIMATIC PDM, AMS, FDT [such as PACTware or Fieldcare]). See SITRANS LR560 (mA/HART) Instruction Manual for more information. • Foundation Fieldbus FF (using 375 handheld Field Communicator or AMS). See SITRANS LR560 (Foundation Fieldbus) Instruction Manual for more information.Once programmed, the graphic Local Display Interface (LDI) can be removed and used to transfer parameters to multiple SITRANS LR560s.](https://usermanual.wiki/Siemens-Canada-Siemens-Milltronics-Process-Instruments/LR560.User-Manual-PA/User-Guide-2312646-Page-12.png)































![Page 38 SITRANS LR560 (PROFIBUS PA) – OPERATING INSTRUCTIONS 7ML19985LT01mmmmmQuick Start: localLevel application exampleThe application is a vessel that takes an average 3 hours (180 minutes) to fill and 3 weeks to empty.Fill rate = 0.08 m/minute [(Low Cal Pt. minus High Cal Pt.) / fastest of fill or empty time]= (15.5 m – 1 m) / 180 min. = 14.5 m /180 min. = 0.08 m/min.Therefore SLOW response rate (0.1 m/minute) can be selected.Quick Start Parameter Setting Description Vessel STEEL Selects vessel construction material.Response Rate SLOW Resets Fill Rate/ Empty Rate to 0.1 m/minute.Units M Sensor measurement units.Operation LEVEL (1) Material level referenced from Low Cal. Point.Low Calibration Point 15.5 Process empty level.High Calibration Point 1.0 Process full level.sensor reference point Level Low Cal. PointHigh Cal. Point15.5 m1.0 mSITRANS LR560](https://usermanual.wiki/Siemens-Canada-Siemens-Milltronics-Process-Instruments/LR560.User-Manual-PA/User-Guide-2312646-Page-44.png)



![Page 42 SITRANS LR560 (PROFIBUS PA) – OPERATING INSTRUCTIONS 7ML19985LT01mmmmmSIMATIC PDMStep 1 – Identification1) Click on Read Data from Device to upload Quick Start parameter settings from the device to the PC/PG and ensure PDM is synchronized with the device.2) Click on NEXT to accept the default values. (Descriptor, Message, and Date fields can be left blank.)Step 2 – Application Type1) Select the application type: steel or concrete vessel. [This changes the setting for Position Detect (2.4.5.2.) to Rising Edge.] 2) Click on NEXT. Note: The layout of the dialog boxes shown may vary according to the resolution set-ting for your computer monitor.Notes: • Selecting either STEEL or CONCRETE does a functional reset (see Master Reset (4.2.) on page 87).• Selecting STEEL changes the setting for Position Detect (2.4.5.2.) to Rising Edge. and for Algorithm (2.4.5.1.) to F.• Selecting CONCRETE changes the setting for Position Detect (2.4.5.2.) to Rising Edge and for Algorithm (2.4.5.1.) to ALF.](https://usermanual.wiki/Siemens-Canada-Siemens-Milltronics-Process-Instruments/LR560.User-Manual-PA/User-Guide-2312646-Page-48.png)


























































![7ML19985LT01 SITRANS LR560 (PROFIBUS PA) – OPERATING INSTRUCTIONS Page 101mmmmmParameters5.2. PROFIBUS Ident NumberIdentifies the device on the network. The Ident Number must match that in the GSD file (the GSD file provides information on the device to the master).6. Security6.1. Remote Access6.1.1. Remote Lockout Enables or disables programming via the network and PDM.6.2. Local Access6.2.1. Write ProtectionPrevents any changes to parameters via PDM or the hand-held programmer. 7. LanguageSelects the language to be used on the LCD.To access Language via SIMATIC PDM:Navigate to Level Meter > Setup > Local Display.OptionsSTD PROFILE Standard Profile (uses generic GSD for 2 AIFB [ident # = 0x9701]* MANUFACTURERManufacturer-specific (uses Siemens EDD and GSD file, which identifies the LR560 [PROFIBUS PA]) [ident # = 0x816B]STD – AIFB 1 ONL. Standard Profile AIFB 1 only (uses generic GSD for 1 AIFB) [ident # = 0x9700]Note: Default settings in the parameter tables are indicated with an asterisk (*) unless explicitly stated.Note: If access control is changed to limit remote access, it can be reset only via the handheld programmer. Options * OFF (Remote operation enabled)ON (Remote operation disabled)Hand-held programmer ValuesRange: 0 to 99992457 (unlock value) Off (enables programming)any other value On (disables programming)Options*ENGLISHDEUTSCHFRANÇAISESPAÑOLCHINESE](https://usermanual.wiki/Siemens-Canada-Siemens-Milltronics-Process-Instruments/LR560.User-Manual-PA/User-Guide-2312646-Page-107.png)























level units(Analog Input)AIFB1Secondary Value 1 [Level](Level Units)Secondary Value 2 [Distance](Sensor Units)Sensor OffsetLevel OffsetHigh Calibration PointHigh Level Point Low Calibration Point Low Level PointCalibration TypePVSV1SV2AIFB2(Analog Input)](https://usermanual.wiki/Siemens-Canada-Siemens-Milltronics-Process-Instruments/LR560.User-Manual-PA/User-Guide-2312646-Page-131.png)




![Page 130 SITRANS LR560 (PROFIBUS PA) – OPERATING INSTRUCTIONS 7ML19985LT01mmmmmF: Communications PAPROFIBUS addressA unique PROFIBUS address identifies each device on the network. To set the PROFIBUS address see Device Address (5.1.) on page 100.Resetting the PROFIBUS address to 126• Via SIMATIC PDM:a) Open the menu Device – Master Reset and click on Reset Address to 126.b) Click on OK: the address will be reset to 126, and if the address lock was on, it will be disabled.• Via the handheld programmer:a) Navigate to Service (4.) > Master Reset (4.2.). (You can enter the numeric value instead of navigating via the Arrow keys.)b) Press RIGHT Arrow to open Edit Mode then scroll down to DEV ADDRESS and press RIGHT Arrow to select it. The address will be reset to 126, and if the address lock was on, it will be disabled.c) Press LEFT Arrow to exit.Operating as a Profile DeviceEvery manufactured PROFIBUS product has a unique PROFIBUS identification number which identifies it to the system. PROFIBUS Profile Standard version 3.01 also defines a Profile Model which can identify a product as a generic profile device on the network. SITRANS LR560 can be identified in one of three ways:Defining the device as Profile-specific as opposed to Manufacturer-specific makes it possible to exchange the device for any other device of the same profile type without changing the GSD file. To set up SITRANS LR560 as a profile device see PROFIBUS Ident Number (5.2.) on page 101.Notes:• It is possible to change the device address via a Class 1 master (for example, a PLC) and lock the device address to prevent further changes.• If this Address Lock is on, the PA address cannot be changed. This lock can be disabled only by performing an Address Reset.Device Identification Profile ModelSTD PROFILE Standard Profile (uses generic GSD for 2 AIFB [ident # = 0x9701]* MANUFACTURER Manufacturer-specific (uses Siemens EDD and GSD file, which identifies the LR560 [PROFIBUS PA]) [ident # = 0x816B]STD – AIFB 1 ONLY Standard Profile AIFB 1 only (uses generic GSD for 1 AIFB) [ident # = 0x9700]](https://usermanual.wiki/Siemens-Canada-Siemens-Milltronics-Process-Instruments/LR560.User-Manual-PA/User-Guide-2312646-Page-136.png)







![Page 138 SITRANS LR560 (PROFIBUS PA) – OPERATING INSTRUCTIONS 7ML19985LT01mmmmmF: Communications PAAcyclic Extended Diagnostics (General Fault Codes)In addition to the extended diagnostics available by cyclic data exchange (shown above), further extended diagnostics are available via acyclic communications. This consists of six bytes. See Diagnosis reply (available cyclically) on page 135 for information on the location of the Extended DiagnosticsNote: Certain fault codes (identified by an asterisk [*] in the table below) will persist until a manual reset has been performed (see Fault Reset (3.2.) on page 82).Acyclic Extended Diagnostics /General Fault Codes LCD display Meaning Corrective Action Byte BitS:0The device was unable to get a measurement within the Fail-safe LOE Timer period. Possible causes: faulty installation, antenna material buildup, foaming/other adverse process conditions, invalid calibration range.• Ensure installation details are correct. • Ensure no antenna material buildup. Clean if necessary. • Adjust process conditions to minimize foam or other adverse conditions. • Correct range calibration.• If fault persists, contact your local Siemens representative.00 S:2Unable to collect profile because of a power condition that is outside the operating range of the device.Repair required. Contact your local Siemens representative. 2 S:3Device is nearing its lifetime limit according to the value set in Maintenance Required Limit.Replacement is recommended.3S:4Device is nearing its lifetime limit according to the value set in Maintenance Demanded Limit.Replacement is recommended.4S:6Sensor is nearing its lifetime limit according to the value set in Maintenance Required Limit.Replacement is recommended.6S:7Sensor is nearing its lifetime limit according to the value set in Maintenance Demanded Limit.Replacement is recommended.7](https://usermanual.wiki/Siemens-Canada-Siemens-Milltronics-Process-Instruments/LR560.User-Manual-PA/User-Guide-2312646-Page-144.png)